The experiment was carried out at Gammasphere
with 92 Ge detectors.
A thin 252Cf source was used allowing the fission partners
to recoil out of the source and be detected by the Rochester parallel plate
avalanche counter (PPAC) array
(CHICO) as shown in Fig. 1.
The source was 70
Ci of 252Cf
electroplated onto a 500
g/cm2 Ni foil with a 20
g/cm2
carbon
cover foil to prevent self-transport of the Cf. Events consisting of 2
particles and at least 3
-rays were collected at
4000 Hz.
The total data set contains
900 million events.
![]() |
The particle detector (CHICO) consists of 2 conical arrays of PPACs
covering a total solid angle of 68
of
sr, although the usable
solid angle for this experiment is
50
due to energy
loss of the fragments in the Ni backing foil. The detector
measures time-of-flight differences of the fission fragments with
500 ps resolution and angular position of the fission axis to
within
in
and
in
. Although PPACs do
not measure the energy of the fission fragments, as silicon
detectors do, they accept large
count rates (up to
1 MHz for CHICO), can be tuned to be insensitive
to the significant
-field (32 times the fission rate),
and are resistant to radiation damage.
The total kinetic energy (TKE) variation of
the nascent fission fragments
as a function of mass division is well known (see for example
Refs. 1,2).
By using this TKE and simple 2-body kinematics, correcting
for energy loss in the backing foil and geometric
variation of the flight path, the time-of-flight
difference gives the mass division and velocities of the
fission partners. The minimum flight path from the source
to the detector is 13 cm, giving a time-of-flight difference
for the strongest channels of about 5 ns. The short flight path
along with the time resolution of the detector limits the
experimental mass resolution to
8 amu
as shown in Fig 2.
![]() |
Since the fission products are recoiling at large velocities,
the
-rays emitted in flight need to be Doppler corrected. As
mentioned above, the velocities of the fission partners are
measured along with the angular position
of the fission axis, allowing the Doppler correction to be performed on an
event-by-event basis. The
-ray resolution after the Doppler correction
is
7 keV for a 1 MeV
-ray, compared to
2 keV
intrinsic detector resolution. The degradation of
resolution after the Doppler correction is due mainly to
the
opening angle of the Gammasphere Ge detectors.